China Tension Meter for Yarn Measurement - High Precision Suppliers and Factory Solutions
DTM200 Model Tension Meter
DTM200 model, a kind of textile testing instruments, is commonly used in textile plant, especially suitable for measuring yarns on line in order to help the process technicians set process parameter as a useful portable tool. Some common applications are listed as the following table.
Technical Specification
| Synthetic fiber plant | for texturing machine |
| Cotton mill | for winding machine |
| Knitting workshop | Used for precise adjustment of yarn feeders on circular knitting machine |
| Optical fiber industry | winding machine |
| Metal wire industry | for wire drawing or winding machine |
Battery
Use screwdriver to dismantle battery cover as above drawing, put in 2 AA batteries (attention: positive/negative), then tight the screw.
ON/OFF
Press POWER key and hold it, when tension meter LCD display shows [.200.], it turns on.
Press POWER key and hold it till tension meter LCD display shows [- - - -], it turns off.
MEASURE
Turn on tension meter, press ZERO key, show 0 on LCD display, press down the handle located in the middle in order that yarn runs on the grooves of ceramic guides which are on both sides and under measuring pin, release the handle to start measure yarn tension.
HOLD ON
During measuring yarn tension, press HOLD key into hold mode to keep data displayed on. Press HOLD key again back to measuring Mode.
ZERO SETTING
Press ZERO key to set 0 values in different positions.
ELECTRONIC DAMP
If yarn tension inconstant when measuring, press DAMP key to get more precise tension data.
Calibrate tension meter (note: the side surface of tension meter must be placed horizontally on a fixed plane during calibrating)
It can be recalibrated by the operator if necessary, it has two steps:
1. Zero value calibration: in measuring mode, press DAMP key for 3 seconds into calibrating mode. Tension meter’s LCD display [C0. ---0] without load, press HOLD key.
2. Full value calibration: after zero value calibration, LCD display [C1. -100]. When tension meter is full load, press the HOLD key to finish full value calibration.
Data Sheet
| Accuracy | ±1% full scale and ±1 digit |
| Units | cN (100cN=1N) |
| Over-range | approx. 10% full scale, without accuracy guarantee |
| Overload protection | 250% of tension range |
| Measuring principle | Strain gauge bridge |
| Measuring roller deflection | 0.5 mm max. |
| Signal process | Digital, 12 bit A/D converter |
| Damp | Adjustable electronic damp |
| Sampling rate | approx. 5 kHz (internal only) |
| Display update rate | approximate 2 times /s |
| Display | 4-digit LCD, 12.7mm high |
| Temperature range | 10-45º C |
| Air humidity | Max. 85% RH |
| Auto power off | Automatically off after aprox. 3 min. of non-use |
| Power supply | Two-Cell Alkaline Battery (2xAA) (about 50 hours of continuous use) |
| Material | Aluminum profile with Aluminum outer casing |
| Dimensions | 230 x 52 x 46 mm (L x W x H); 350 x 270x 80 mm (L x W x H) |
| Net weight (gross) | approximate 500 g (1020 g) |
Frequently Asked Questions
What is the DTM200 Model Tension Meter used for?
It is a textile testing instrument commonly used in textile plants to measure yarns on line, helping process technicians set precise process parameters.
What type of battery does the DTM200 use?
It operates on two AA Alkaline batteries, providing approximately 50 hours of continuous use.
How do you turn the tension meter on and off?
Press and hold the POWER key. When the LCD display shows [.200.], it turns on. To turn it off, press and hold the POWER key until the display shows [- - - -].
What is the measurement accuracy of the DTM200?
The tension meter features an accuracy of ±1% full scale and ±1 digit.
How do I calibrate the DTM200 Model Tension Meter?
Place the device horizontally on a fixed plane. First, perform Zero value calibration by pressing the DAMP key for 3 seconds in measuring mode, then press HOLD when [C0. ---0] is displayed. Next, perform Full value calibration by pressing the HOLD key when [C1. -100] is shown under full load.
What should I do if the yarn tension reading is inconstant?
If the yarn tension is inconstant during measurement, press the DAMP key to activate the adjustable electronic damp for more precise tension data.